ASCY Home Page

For Screening Staff

Note: An online step-by-step training program is available through the Thunder Bay District Health Unit website.

Testing Hints for DPOAE

  • Swaddle baby and position on side for best access to ear
  • Best to test 24 hours after birth to prevent false positive due to vernix
  • If infant fails, retest once more before 6-week period
  • Testing can be completed while baby is nursing or feeding
  • If infant is fussy or crying, offer pacifier

Troubleshooting Hints for DPOAE

  • Conduct a visual inspection of the probe before each use
  • Clean the probe if debris is noted
    • remove probe tip
    • remove microphone
    • remove clear tubing from probe body
    • thread cleaning wire through all 3 metal tubes on probe body until the tubes are clear of debris
    • if probe tip is clogged, use a new probe tip

Error Messages for DPOAE

PROBE IN EAR? REFIT AND RETRY OR OCCULED PROBE

  • clean probe
  • verify a secure connection of all cables, tubing and mirophone
  • verify that no pins are bent or broken on the round end of the probe cable connector
  • verify that there is no damage to the round probe connection on the AuDX box
  • verify that the clear tubing has no holes, tears or condensation, replace if necessary
  • retest baby - if error continues, contact the Infant Hearing Program @ 905-385-7927 ext 227.

TOO NOISY/HIGH ARTIFACTS

  • remove probe from ear
  • verify proper use of probe tip (size) for baby's ear
  • re-insert probe into baby's ear, ensuring a secure seal
  • verify a quiet testing environment, including a quiet baby

NO LOADER RELOAD SOFTWARE or MEMORY ERROR RELOAD SOFTWARE

Contact the Infant Hearing Program 905-385-7927 ext. 227

IF ERROR MESSAGES CONTINUE -
TEST YOURSELF TO SEE IF PROBE IS WORKING PROPERLY
(FREQUENCY TONES CAN BE HEARD)